§360c. Classification of devices intended for human use
(a) Classes of devices
(1) There are established the following classes of devices intended for human use:
(A)
(i) A device for which the controls authorized by or under section 351, 352, 360, 360f, 360h, 360i, or 360j of this title or any combination of such sections are sufficient to provide reasonable assurance of the safety and effectiveness of the device.
(ii) A device for which insufficient information exists to determine that the controls referred to in clause (i) are sufficient to provide reasonable assurance of the safety and effectiveness of the device or to establish special controls to provide such assurance, but because it-
(I) is not purported or represented to be for a use in supporting or sustaining human life or for a use which is of substantial importance in preventing impairment of human health, and
(II) does not present a potential unreasonable risk of illness or injury,
is to be regulated by the controls referred to in clause (i).
(B)
(C)
(i) it (I) cannot be classified as a class I device because insufficient information exists to determine that the application of general controls are sufficient to provide reasonable assurance of the safety and effectiveness of the device, and (II) cannot be classified as a class II device because insufficient information exists to determine that the special controls described in subparagraph (B) would provide reasonable assurance of its safety and effectiveness, and
(ii)(I) is purported or represented to be for a use in supporting or sustaining human life or for a use which is of substantial importance in preventing impairment of human health, or
(II) presents a potential unreasonable risk of illness or injury,
is to be subject, in accordance with section 360e of this title, to premarket approval to provide reasonable assurance of its safety and effectiveness.
If there is not sufficient information to establish a performance standard for a device to provide reasonable assurance of its safety and effectiveness, the Secretary may conduct such activities as may be necessary to develop or obtain such information.
(2) For purposes of this section and sections 360d and 360e of this title, the safety and effectiveness of a device are to be determined-
(A) with respect to the persons for whose use the device is represented or intended,
(B) with respect to the conditions of use prescribed, recommended, or suggested in the labeling of the device, and
(C) weighing any probable benefit to health from the use of the device against any probable risk of injury or illness from such use.
(3)(A) Except as authorized by subparagraph (B), the effectiveness of a device is, for purposes of this section and sections 360d and 360e of this title, to be determined, in accordance with regulations promulgated by the Secretary, on the basis of well-controlled investigations, including clinical investigations where appropriate, by experts qualified by training and experience to evaluate the effectiveness of the device, from which investigations it can fairly and responsibly be concluded by qualified experts that the device will have the effect it purports or is represented to have under the conditions of use prescribed, recommended, or suggested in the labeling of the device.
(B) If the Secretary determines that there exists valid scientific evidence (other than evidence derived from investigations described in subparagraph (A))-
(i) which is sufficient to determine the effectiveness of a device, and
(ii) from which it can fairly and responsibly be concluded by qualified experts that the device will have the effect it purports or is represented to have under the conditions of use prescribed, recommended, or suggested in the labeling of the device,
then, for purposes of this section and sections 360d and 360e of this title, the Secretary may authorize the effectiveness of the device to be determined on the basis of such evidence.
(b) Classification panels
(1) For purposes of-
(A) determining which devices intended for human use should be subject to the requirements of general controls, performance standards, or premarket approval, and
(B) providing notice to the manufacturers and importers of such devices to enable them to prepare for the application of such requirements to devices manufactured or imported by them,
the Secretary shall classify all such devices (other than devices classified by subsection (f) of this section) into the classes established by subsection (a) of this section. For the purpose of securing recommendations with respect to the classification of devices, the Secretary shall establish panels of experts or use panels of experts established before May 28, 1976, or both. Section 14 of the Federal Advisory Committee Act shall not apply to the duration of a panel established under this paragraph.
(2) The Secretary shall appoint to each panel established under paragraph (1) persons who are qualified by training and experience to evaluate the safety and effectiveness of the devices to be referred to the panel and who, to the extent feasible, possess skill in the use of, or experience in the development, manufacture, or utilization of, such devices. The Secretary shall make appointments to each panel so that each panel shall consist of members with adequately diversified expertise in such fields as clinical and administrative medicine, engineering, biological and physical sciences, and other related professions. In addition, each panel shall include as nonvoting members a representative of consumer interests and a representative of interests of the device manufacturing industry. Scientific, trade, and consumer organizations shall be afforded an opportunity to nominate individuals for appointment to the panels. No individual who is in the regular full-time employ of the United States and engaged in the administration of this chapter may be a member of any panel. The Secretary shall designate one of the members of each panel to serve as chairman thereof.
(3) Panel members (other than officers or employees of the United States), while attending meetings or conferences of a panel or otherwise engaged in its business, shall be entitled to receive compensation at rates to be fixed by the Secretary, but not at rates exceeding the daily equivalent of the rate in effect for grade GS–18 of the General Schedule, for each day so engaged, including traveltime; and while so serving away from their homes or regular places of business each member may be allowed travel expenses (including per diem in lieu of subsistence) as authorized by section 5703 of title 5, for persons in the Government service employed intermittently.
(4) The Secretary shall furnish each panel with adequate clerical and other necessary assistance.
(c) Classification panel organization and operation
(1) The Secretary shall organize the panels according to the various fields of clinical medicine and fundamental sciences in which devices intended for human use are used. The Secretary shall refer a device to be classified under this section to an appropriate panel established or authorized to be used under subsection (b) of this section for its review and for its recommendation respecting the classification of the device. The Secretary shall by regulation prescribe the procedure to be followed by the panels in making their reviews and recommendations. In making their reviews of devices, the panels, to the maximum extent practicable, shall provide an opportunity for interested persons to submit data and views on the classification of the devices.
(2)(A) Upon completion of a panel's review of a device referred to it under paragraph (1), the panel shall, subject to subparagraphs (B) and (C), submit to the Secretary its recommendation for the classification of the device. Any such recommendation shall (i) contain (I) a summary of the reasons for the recommendation, (II) a summary of the data upon which the recommendation is based, and (III) an identification of the risks to health (if any) presented by the device with respect to which the recommendation is made, and (ii) to the extent practicable, include a recommendation for the assignment of a priority for the application of the requirements of section 360d or 360e of this title to a device recommended to be classified in class II or class III.
(B) A recommendation of a panel for the classification of a device in class I shall include a recommendation as to whether the device should be exempted from the requirements of section 360, 360i, or 360j(f) of this title.
(C) In the case of a device which has been referred under paragraph (1) to a panel, and which-
(i) is intended to be implanted in the human body or is purported or represented to be for a use in supporting or sustaining human life, and
(ii)(I) has been introduced or delivered for introduction into interstate commerce for commercial distribution before May 28, 1976, or
(II) is within a type of device which was so introduced or delivered before such date and is substantially equivalent to another device within that type,
such panel shall recommend to the Secretary that the device be classified in class III unless the panel determines that classification of the device in such class is not necessary to provide reasonable assurance of its safety and effectiveness. If a panel does not recommend that such a device be classified in class III, it shall in its recommendation to the Secretary for the classification of the device set forth the reasons for not recommending classification of the device in such class.
(3) The panels shall submit to the Secretary within one year of the date funds are first appropriated for the implementation of this section their recommendations respecting all devices of a type introduced or delivered for introduction into interstate commerce for commercial distribution before May 28, 1976.
(d) Panel recommendation; publication; priorities
(1) Upon receipt of a recommendation from a panel respecting a device, the Secretary shall publish in the Federal Register the panel's recommendation and a proposed regulation classifying such device and shall provide interested persons an opportunity to submit comments on such recommendation and the proposed regulation. After reviewing such comments, the Secretary shall, subject to paragraph (2), by regulation classify such device.
(2)(A) A regulation under paragraph (1) classifying a device in class I shall prescribe which, if any, of the requirements of section 360, 360i, or 360j(f) of this title shall not apply to the device. A regulation which makes a requirement of section 360, 360i, or 360j(f) of this title inapplicable to a device shall be accompanied by a statement of the reasons of the Secretary for making such requirement inapplicable.
(B) A device described in subsection (c)(2)(C) of this section shall be classified in class III unless the Secretary determines that classification of the device in such class is not necessary to provide reasonable assurance of its safety and effectiveness. A proposed regulation under paragraph (1) classifying such a device in a class other than class III shall be accompanied by a full statement of the reasons of the Secretary (and supporting documentation and data) for not classifying such device in such class and an identification of the risks to health (if any) presented by such device.
(3) In the case of devices classified in class II and devices classified under this subsection in class III and described in section 360e(b)(1) of this title the Secretary may establish priorities which, in his discretion, shall be used in applying sections 360d and 360e of this title, as appropriate, to such devices.
(e) Classification changes
(1) Based on new information respecting a device, the Secretary may, upon his own initiative or upon petition of an interested person, by regulation (A) change such device's classification, and (B) revoke, because of the change in classification, any regulation or requirement in effect under section 360d or 360e of this title with respect to such device. In the promulgation of such a regulation respecting a device's classification, the Secretary may secure from the panel to which the device was last referred pursuant to subsection (c) of this section a recommendation respecting the proposed change in the device's classification and shall publish in the Federal Register any recommendation submitted to the Secretary by the panel respecting such change. A regulation under this subsection changing the classification of a device from class III to class II may provide that such classification shall not take effect until the effective date of a performance standard established under section 360d of this title for such device.
(2) By regulation promulgated under paragraph (1), the Secretary may change the classification of a device from class III-
(A) to class II if the Secretary determines that special controls would provide reasonable assurance of the safety and effectiveness of the device and that general controls would not provide reasonable assurance of the safety and effectiveness of the device, or
(B) to class I if the Secretary determines that general controls would provide reasonable assurance of the safety and effectiveness of the device.
(f) Initial classification and reclassification of certain devices
(1) Any device intended for human use which was not introduced or delivered for introduction into interstate commerce for commercial distribution before May 28, 1976, is classified in class III unless-
(A) the device-
(i) is within a type of device (I) which was introduced or delivered for introduction into interstate commerce for commercial distribution before such date and which is to be classified pursuant to subsection (b) of this section, or (II) which was not so introduced or delivered before such date and has been classified in class I or II, and
(ii) is substantially equivalent to another device within such type, or
(B) the Secretary in response to a petition submitted under paragraph (2) has classified such device in class I or II.
A device classified in class III under this paragraph shall be classified in that class until the effective date of an order of the Secretary under paragraph (2) classifying the device in class I or II.
(2)(A) The Secretary may initiate the reclassification of a device classified into class III under paragraph (1) of this subsection or the manufacturer or importer of a device classified under paragraph (1) may petition the Secretary (in such form and manner as he shall prescribe) for the issuance of an order classifying the device in class I or class II. Within thirty days of the filing of such a petition, the Secretary shall notify the petitioner of any deficiencies in the petition which prevent the Secretary from making a decision on the petition.
(B)(i) Upon determining that a petition does not contain any deficiency which prevents the Secretary from making a decision on the petition, the Secretary may for good cause shown refer the petition to an appropriate panel established or authorized to be used under subsection (b) of this section. A panel to which such a petition has been referred shall not later than ninety days after the referral of the petition make a recommendation to the Secretary respecting approval or denial of the petition. Any such recommendation shall contain (I) a summary of the reasons for the recommendation, (II) a summary of the data upon which the recommendation is based, and (III) an identification of the risks to health (if any) presented by the device with respect to which the petition was filed. In the case of a petition for a device which is intended to be implanted in the human body or which is purported or represented to be for a use in supporting or sustaining human life, the panel shall recommend that the petition be denied unless the panel determines that the classification in class III of the device is not necessary to provide reasonable assurance of its safety and effectiveness. If the panel recommends that such petition be approved, it shall in its recommendation to the Secretary set forth its reasons for such recommendation.
(ii) The requirements of paragraphs (1) and (2) of subsection (c) of this section (relating to opportunities for submission of data and views and recommendations respecting priorities and exemptions from sections 360, 360i, and 360j(f) of this title) shall apply with respect to consideration by panels of petitions submitted under subparagraph (A).
(C)(i) Within ninety days from the date the Secretary receives the recommendation of a panel respecting a petition (but not later than 210 days after the filing of such petition) the Secretary shall by order deny or approve the petition. If the Secretary approves the petition, the Secretary shall order the classification of the device into class I or class II in accordance with the criteria prescribed by subsection (a)(1)(A) or (a)(1)(B) of this section. In the case of a petition for a device which is intended to be implanted in the human body or which is purported or represented to be for a use in supporting or sustaining human life, the Secretary shall deny the petition unless the Secretary determines that the classification in class III of the device is not necessary to provide reasonable assurance of its safety and effectiveness. An order approving such petition shall be accompanied by a full statement of the reasons of the Secretary (and supporting documentation and data) for approving the petition and an identification of the risks to health (if any) presented by the device to which such order applies.
(ii) The requirements of paragraphs (1) and (2)(A) of subsection (d) of this section (relating to publication of recommendations, opportunity for submission of comments, and exemption from sections 360, 360i, and 360j(f) of this title) shall apply with respect to action by the Secretary on petitions submitted under subparagraph (A).
(3) If a manufacturer reports to the Secretary under section 360(k) of this title that a device is substantially equivalent to another device-
(A) which the Secretary has classified as a class III device under subsection (b) of this section,
(B) which was introduced or delivered for introduction into interstate commerce for commercial distribution before December 1, 1990, and
(C) for which no final regulation requiring premarket approval has been promulgated under section 360e(b) of this title,
the manufacturer shall certify to the Secretary that the manufacturer has conducted a reasonable search of all information known or otherwise available to the manufacturer respecting such other device and has included in the report under section 360(k) of this title a summary of and a citation to all adverse safety and effectiveness data respecting such other device and respecting the device for which the section 360(k) report is being made and which has not been submitted to the Secretary under section 360i of this title. The Secretary may require the manufacturer to submit the adverse safety and effectiveness data described in the report.
(g) Information
Within sixty days of the receipt of a written request of any person for information respecting the class in which a device has been classified or the requirements applicable to a device under this chapter, the Secretary shall provide such person a written statement of the classification (if any) of such device and the requirements of this chapter applicable to the device.
(h) Definitions
For purposes of this section and sections 351, 360, 360d, 360e, 360f, 360i, and 360j of this title
(1) a reference to "general controls" is a reference to the controls authorized by or under sections 351, 352, 360, 360f, 360h, 360i, and 360j of this title,
(2) a reference to "class I", "class II", or "class III" is a reference to a class of medical devices described in subparagraph (A), (B), or (C) of subsection (a)(1) of this section, and
(3) a reference to a "panel under section 360c of this title" is a reference to a panel established or authorized to be used under this section.
(i) Substantial equivalence
(1)(A) For purposes of determinations of substantial equivalence under subsection (f) of this section and section 360j(l) of this title, the term "substantially equivalent" or "substantial equivalence" means, with respect to a device being compared to a predicate device, that the device has the same intended use as the predicate device and that the Secretary by order has found that the device-
(i) has the same technological characteristics as the predicate device, or
(ii)(I) has different technological characteristics and the information submitted that the device is substantially equivalent to the predicate device contains information, including clinical data if deemed necessary by the Secretary, that demonstrates that the device is as safe and effective as a legally marketed device, and (II) does not raise different questions of safety and efficacy than the predicate device.
(B) For purposes of subparagraph (A), the term "different technological characteristics" means, with respect to a device being compared to a predicate device, that there is a significant change in the materials, design, energy source, or other features of the device from those of the predicate device.
(2) A device may not be found to be substantially equivalent to a predicate device that has been removed from the market at the initiative of the Secretary or that has been determined to be misbranded or adulterated by a judicial order.
(3)(A) As part of a submission under section 360(k) of this title respecting a device, the person required to file a premarket notification under such section shall provide an adequate summary of any information respecting safety and effectiveness or state that such information will be made available upon request by any person.
(B) Any summary under subparagraph (A) respecting a device shall contain detailed information regarding data concerning adverse health effects and shall be made available to the public by the Secretary within 30 days of the issuance of a determination that such device is substantially equivalent to another device.
(June 25, 1938, ch. 675, §513, as added May 28, 1976,
References in Text
Section 14 of the Federal Advisory Committee Act, referred to in subsec. (b)(1), is section 14 of
Amendments
1993-Subsec. (b)(3).
1992-Subsec. (f)(3).
1990-Subsec. (a)(1)(A)(ii).
Subsec. (a)(1)(B).
Subsec. (a)(1)(C)(i).
Subsec. (e).
Subsec. (f).
Subsec. (f)(2)(A).
Subsec. (f)(2)(B)(i).
Subsec. (f)(3).
Subsec. (i).
Short Title of 1976 Amendment
Regulations
Section 12(b) of
Daily Wear Soft or Daily Wear Nonhydrophilic Plastic Contact Lenses
Section 4(b)(3) of
"(A) Notwithstanding section 520(l)(5) of the Federal Food, Drug, and Cosmetic Act [21 U.S.C. 360j(l)(5)], the Secretary of Health and Human Services shall not retain any daily wear soft or daily wear nonhydrophilic plastic contact lens in class III under such Act [this chapter] unless the Secretary finds that it meets the criteria set forth in section 513(a)(1)(C) of such Act [21 U.S.C. 360c(a)(1)(C)]. The finding and the grounds for the finding shall be published in the Federal Register. For any such lens, the Secretary shall make the determination respecting reclassification required in section 520(l)(5)(B) of such Act within 24 months of the date of the enactment of this paragraph [Nov. 28, 1990].
"(B) The Secretary of Health and Human Services may by notice published in the Federal Register extend the two-year period prescribed by subparagraph (A) for a lens for an additional period not to exceed one year.
"(C)(i) Before classifying a lens in class II pursuant to subparagraph (A), the Secretary of Health and Human Services shall pursuant to section 513(a)(1)(B) of such Act assure that appropriate regulatory safeguards are in effect which provide reasonable assurance of the safety and effectiveness of such lens, including clinical and preclinical data if deemed necessary by the Secretary.
"(ii) Prior to classifying a lens in class I pursuant to subparagraph (A), the Secretary shall assure that appropriate regulatory safeguards are in effect which provide reasonable assurance of the safety and effectiveness of such lens, including clinical and preclinical data if deemed necessary by the Secretary.
"(D) Notwithstanding section 520(l)(5) of such Act, if the Secretary of Health and Human Services has not made the finding and published the finding required by subparagraph (A) within 36 months of the date of the enactment of this subparagraph [Nov. 28, 1990], the Secretary shall issue an order placing the lens in class II.
"(E) Any person adversely affected by a final regulation under this paragraph revising the classification of a lens may challenge the revision of the classification of such lens only by filing a petition under section 513(e) for a classification change."