21 USC 360i: Records and reports on devices
Result 1 of 1
   
 
21 USC 360i: Records and reports on devices Text contains those laws in effect on January 4, 1995
From Title 21-FOOD AND DRUGSCHAPTER 8-NARCOTIC FARMSSUBCHAPTER V-DRUGS AND DEVICESPart A-Drugs and Devices

§360i. Records and reports on devices

(a) General rule

Every person who is a manufacturer, importer, or distributor of a device intended for human use shall establish and maintain such records, make such reports, and provide such information, as the Secretary may by regulation reasonably require to assure that such device is not adulterated or misbranded and to otherwise assure its safety and effectiveness. Regulations prescribed under the preceding sentence-

(1) shall require a device manufacturer or importer to report to the Secretary whenever the manufacturer or importer receives or otherwise becomes aware of information that reasonably suggests that one of its marketed devices-

(A) may have caused or contributed to a death or serious injury, or

(B) has malfunctioned and that such device or a similar device marketed by the manufacturer or importer would be likely to cause or contribute to a death or serious injury if the malfunction were to recur;


(2) shall define the term "serious injury" to mean an injury that-

(A) is life threatening,

(B) results in permanent impairment of a body function or permanent damage to a body structure, or

(C) necessitates medical or surgical intervention to preclude permanent impairment of a body function or permanent damage to a body structure;


(3) shall require reporting of other significant adverse device experiences as determined by the Secretary to be necessary to be reported;

(4) shall not impose requirements unduly burdensome to a device manufacturer, importer, or distributor taking into account his cost of complying with such requirements and the need for the protection of the public health and the implementation of this chapter;

(5) which prescribe the procedure for making requests for reports or information shall require that each request made under such regulations for submission of a report or information to the Secretary state the reason or purpose for such request and identify to the fullest extent practicable such report or information;

(6) which require submission of a report or information to the Secretary shall state the reason or purpose for the submission of such report or information and identify to the fullest extent practicable such report or information;

(7) may not require that the identity of any patient be disclosed in records, reports, or information required under this subsection unless required for the medical welfare of an individual, to determine the safety or effectiveness of a device, or to verify a record, report, or information submitted under this chapter;

(8) may not require a manufacturer, importer, or distributor of a class I device to-

(A) maintain for such a device records respecting information not in the possession of the manufacturer, importer, or distributor, or

(B) to submit for such a device to the Secretary any report or information-

(i) not in the possession of the manufacturer, importer, or distributor, or

(ii) on a periodic basis,


unless such report or information is necessary to determine if the device should be reclassified or if the device is adulterated or misbranded; and


(9) shall require distributors who submit such reports to submit copies of the reports to the manufacturer of the device for which the report was made.


In prescribing such regulations, the Secretary shall have due regard for the professional ethics of the medical profession and the interests of patients. The prohibitions of paragraph (7) of this subsection continue to apply to records, reports, and information concerning any individual who has been a patient, irrespective of whether or when he ceases to be a patient.

(b) User reports

(1)(A) Whenever a device user facility receives or otherwise becomes aware of information that reasonably suggests that a device has or may have caused or contributed to the death of a patient of the facility, the facility shall, as soon as practicable but not later than 10 working days after becoming aware of the information, report the information to the Secretary and, if the identity of the manufacturer is known, to the manufacturer of the device. In the case of deaths, the Secretary may by regulation prescribe a shorter period for the reporting of such information.

(B) Whenever a device user facility receives or otherwise becomes aware of-

(i) information that reasonably suggests that a device has or may have caused or contributed to the serious illness of, or serious injury to, a patient of the facility, or

(ii) other significant adverse device experiences as determined by the Secretary by regulation to be necessary to be reported,


the facility shall, as soon as practicable but not later than 10 working days after becoming aware of the information, report the information to the manufacturer of the device or to the Secretary if the identity of the manufacturer is not known.

(C) Each device user facility shall submit to the Secretary on a semi-annual basis a summary of the reports made under subparagraphs (A) and (B). Such summary shall be submitted on January 1 and July 1 of each year. The summary shall be in such form and contain such information from such reports as the Secretary may require and shall include-

(i) sufficient information to identify the facility which made the reports for which the summary is submitted,

(ii) in the case of any product which was the subject of a report, the product name, serial number, and model number,

(iii) the name and the address of the manufacturer of such device, and

(iv) a brief description of the event reported to the manufacturer.


The Secretary may by regulation alter the frequency and timing of reports required by this subparagraph.

(D) For purposes of subparagraphs (A), (B), and (C), a device user facility shall be treated as having received or otherwise become aware of information with respect to a device of that facility when medical personnel who are employed by or otherwise formally affiliated with the facility receive or otherwise become aware of information with respect to that device in the course of their duties.

(2) The Secretary may not disclose the identity of a device user facility which makes a report under paragraph (1) except in connection with-

(A) an action brought to enforce section 331(q) of this title,

(B) a communication to a manufacturer of a device which is the subject of a report under paragraph (1), or

(C) a disclosure required under subsection (a) of this section.


This paragraph does not prohibit the Secretary from disclosing the identity of a device user facility making a report under paragraph (1) or any information in such a report to employees of the Department of Health and Human Services, to the Department of Justice, or to the duly authorized committees and subcommittees of the Congress.

(3) No report made under paragraph (1) by-

(A) a device user facility,

(B) an individual who is employed by or otherwise formally affiliated with such a facility, or

(C) a physician who is not required to make such a report,


shall be admissible into evidence or otherwise used in any civil action involving private parties unless the facility, individual, or physician who made the report had knowledge of the falsity of the information contained in the report.

(4) A report made under paragraph (1) does not affect any obligation of a manufacturer who receives the report to file a report as required under subsection (a) of this section.

(5) For purposes of this subsection:

(A) The term "device user facility" means a hospital, ambulatory surgical facility, nursing home, or outpatient treatment facility which is not a physician's office. The Secretary may by regulation include an outpatient diagnostic facility which is not a physician's office in such term.

(B) The terms "serious illness" and "serious injury" mean illness or injury, respectively, that-

(i) is life threatening,

(ii) results in permanent impairment of a body function or permanent damage to a body structure, or

(iii) necessitates medical or surgical intervention to preclude permanent impairment of a body function or permanent damage to a body structure.

(c) Persons exempt

Subsection (a) of this section shall not apply to-

(1) any practitioner who is licensed by law to prescribe or administer devices intended for use in humans and who manufactures or imports devices solely for use in the course of his professional practice;

(2) any person who manufactures or imports devices intended for use in humans solely for such person's use in research or teaching and not for sale (including any person who uses a device under an exemption granted under section 360j(g) of this title); and

(3) any other class of persons as the Secretary may by regulation exempt from subsection (a) of this section upon a finding that compliance with the requirements of such subsection by such class with respect to a device is not necessary to (A) assure that a device is not adulterated or misbranded or (B) otherwise to assure its safety and effectiveness.

(d) Certification

Each manufacturer, importer, and distributor required to make reports under subsection (a) of this section shall submit to the Secretary annually a statement certifying that-

(1) the manufacturer, importer, or distributor did file a certain number of such reports, or

(2) the manufacturer, importer, or distributor did not file any report under subsection (a) of this section.

(e) Device tracking

Every person who registers under section 360 of this title and is engaged in the manufacture of-

(1) a device the failure of which would be reasonably likely to have serious adverse health consequences and which is (A) a permanently implantable device, or (B) a life sustaining or life supporting device used outside a device user facility, or

(2) any other device which the Secretary may designate,


shall adopt a method of device tracking.

(f) Reports of removals and corrections

(1) Except as provided in paragraph (2), the Secretary shall by regulation require a manufacturer, importer, or distributor of a device to report promptly to the Secretary any correction or removal of a device undertaken by such manufacturer, importer, or distributor if the removal or correction was undertaken-

(A) to reduce a risk to health posed by the device, or

(B) to remedy a violation of this chapter caused by the device which may present a risk to health.


A manufacturer, importer, or distributor of a device who undertakes a correction or removal of a device which is not required to be reported under this paragraph shall keep a record of such correction or removal.

(2) No report of the corrective action or removal of a device may be required under paragraph (1) if a report of the corrective action or removal is required and has been submitted under subsection (a) of this section.

(3) For purposes of paragraphs (1) and (2), the terms "correction" and "removal" do not include routine servicing.

(June 25, 1938, ch. 675, §519, as added May 28, 1976, Pub. L. 94–295, §2, 90 Stat. 564 ; amended Nov. 28, 1990, Pub. L. 101–629, §§2(a), 3(a)(1), (b)(1), 7, 104 Stat. 4511 , 4513, 4514, 4520; June 16, 1992, Pub. L. 102–300, §5(a), 106 Stat. 239 ; Aug. 13, 1993, Pub. L. 103–80, §3(u), 107 Stat. 778 .)

Amendments

1993-Subsec. (a). Pub. L. 103–80 substituted "paragraph (7)" for "paragraph (4)" in last sentence.

1992-Subsec. (a). Pub. L. 102–300, §5(a)(1), added pars. (1) to (3) and redesignated former pars. (1) to (6) as (4) to (9), respectively.

Subsec. (b)(1)(A). Pub. L. 102–300, §5(a)(2)(A), substituted "a device has or may have" for "there is a probability that a device has".

Subsec. (b)(1)(B). Pub. L. 102–300, §5(a)(2)(A), (B), substituted "a device has or may have" for "there is a probability that a device has", designated existing provisions as cl. (i), and added cl. (ii).

Subsec. (b)(5)(B)(iii). Pub. L. 102–300, §5(a)(2)(C), struck out "immediate" before "medical".

1990-Subsec. (a)(6). Pub. L. 101–629, §3(a)(1), added par. (6).

Subsecs. (b), (c). Pub. L. 101–629, §2(a), added subsec. (b) and redesignated former subsec. (b) as (c).

Subsecs. (d), (e). Pub. L. 101–629, §3(b)(1), added subsecs. (d) and (e).

Subsec. (f). Pub. L. 101–629, §7, added subsec. (f).

Effective Date of 1992 Amendment

Section 2(b) of Pub. L. 102–300 provided that: "The amendments made by subsection (a) [amending sections 3(b)(3) and 3(c) of Pub. L. 101–629, set out as notes below] shall take effect as of May 27, 1992 and any rule to implement section 519(e) of the Federal Food, Drug, and Cosmetic Act [21 U.S.C. 360i(e)] proposed under section 3(c)(2) of the Safe Medical Devices Act of 1990 [Pub. L. 101–629, set out as a note below] shall revert to its proposed status as of such date."

Section 5(b) of Pub. L. 102–300 provided that: "The amendments made by subsection (a) [amending this section] shall take effect-

"(1) 1 year after the date of the enactment of this Act [June 16, 1992]; or

"(2) on the effective date of regulations of the Secretary to implement such amendments,

whichever occurs first."

Effective Date of 1990 Amendment

Section 2(c) of Pub. L. 101–629 provided that: "Section 519(b) of the Federal Food, Drug, and Cosmetic Act [21 U.S.C. 360i(b)], as added by the amendment made by subsection (a), shall take effect-

"(1) upon the effective date of regulations promulgated under subsection (b) [set out below], or

"(2) upon the expiration of 12 months from the date of the enactment of this Act [Nov. 28, 1990],

whichever occurs first."

Section 3(a)(2) of Pub. L. 101–629 provided that: "Section 519(a)(6) [21 U.S.C. 360i(a)(6)], as added by the amendment made by paragraph (1), shall take effect upon the effective date of final regulations under subsection (c) [set out below]."

Section 3(b)(3) of Pub. L. 101–629, as amended by Pub. L. 102–300, §2(a)(1), June 16, 1992, 106 Stat. 238 , provided that: "Section 519(e) [21 U.S.C. 360i(e)], as added by the amendment made by paragraph (1), shall take effect upon the expiration of 9 months after the issuance of final regulations under subsection (c) [set out below]."

[For effective date of amendment by Pub. L. 102–300, see section 2(b) of Pub. L. 102–300, set out above as an Effective Date of 1992 Amendment note.]

Regulations

Section 2(b) of Pub. L. 101–629 provided that: "The Secretary of Health and Human Services shall promulgate regulations to implement section 519(b) of the Federal Food, Drug, and Cosmetic Act [21 U.S.C. 360i(b)], as added by the amendment made by subsection (a) (including a definition of the summary required by paragraph (1)(C) of such section) not later than 12 months after the date of enactment of this Act [Nov. 28, 1990]. In promulgating the regulations, the Secretary shall minimize the administrative burdens on device user facilities consistent with the need to assure adequate information."

Section 3(c) of Pub. L. 101–629, as amended by Pub. L. 102–300, §2(a)(2), (3), June 16, 1992, 106 Stat. 238 , provided that:

"(1)(A) Not later than 9 months after the date of the enactment of this Act [Nov. 28, 1990], the Secretary of Health and Human Services shall issue proposed regulations-

"(i) to require distributors of devices to establish and maintain records and to make reports (including reports required by part 803 of title 21 of the Code of Federal Regulations) under section 519(a)(6) of the Federal Food, Drug, and Cosmetic Act [21 U.S.C. 360i(a)(6)], and

"(ii) to implement section 519(e) of such Act.

The Secretary may exempt from regulations described in clause (i) classes of distributors of class I and class II devices from whom reports are not necessary for the protection of the public health.

"(B) Regulations under subparagraph (A) shall-

"(i) require appropriate methods for maintenance of records to ensure that patients who receive devices can be provided the notification required by such Act [this chapter],

"(ii) require that manufacturers adopt effective methods of tracking devices,

"(iii) take into account the position of distributors in the device distribution process, and

"(iv) include such other requirements as the Secretary deems necessary for the adoption of an effective user tracking program under section 519(e) of such Act.

"(2) Not later than 18 months after the date of the enactment of this Act, the Secretary shall issue final regulations to implement sections [sic] 519(a)(6) of the Federal Food, Drug, and Cosmetic Act. If the Secretary does not promulgate such final regulations upon the expiration of such 18 months, the Congress finds that there is good cause for the proposed regulations to be considered as the final regulations without response to comment because the implementation of sections [sic] 519(a)(6) of such Act is essential to protect the health of patients who use such devices. Consequently, in such event, the proposed regulations issued under paragraph (1) shall become final regulations as of the expiration of such 18 months. There shall be promptly published in the Federal Register notice of the new status of the proposed regulations.

"(3) Not later than November 28, 1992, the Secretary shall issue final regulations to implement section 519(e) of the Federal Food, Drug, and Cosmetic Act. If the Secretary does not promulgate such final regulations by November 28, 1992, the Congress finds that there is good cause for the proposed regulations to be considered as the final regulations without response to comment because the implementation of section 519(e) of such Act is essential to protect the health of patients who use devices. In such event, the proposed regulations issued under paragraph (1) shall become the issued final regulations on November 29, 1992. There shall be promptly published in the Federal Register notice of the new status of the proposed regulations."

[For effective date of amendment by Pub. L. 102–300, see section 2(b) of Pub. L. 102–300, set out above as an Effective Date of 1992 Amendment note.]

Information Concerning Reporting Requirements for Device User Facilities

Section 2(d) of Pub. L. 101–629 directed Secretary of Health and Human Services, during the 18-month period beginning on Nov. 28, 1990, to inform device user facilities (as defined in 21 U.S.C. 360i(b)(5)(A)) and manufacturers and distributors of devices respecting the requirements of 21 U.S.C. 360i(b), and, to the extent practicable, provide persons subject to such requirements assistance in the form of publications regarding such requirements.

Study of Reporting Requirements; Compliance by Device User Facilities; Actions by Manufacturers; Cost Effectiveness; Recommendations

Section 2(e) of Pub. L. 101–629 directed Comptroller General of the United States, not more than 36 months after Nov. 28, 1990, to conduct a study of compliance by device user facilities with the requirements of 21 U.S.C. 360i(b), actions taken by manufacturers of devices in response to reports made to them, cost effectiveness of such requirements and their implementation, and any recommendations for improvements to such requirements, with Comptroller General to complete the study and submit a report on the study not later than 45 months from Nov. 28, 1990, to appropriate committees of Congress.

Report to Congress on Reporting Requirements for Device User Facilities

Section 2(f) of Pub. L. 101–629 directed Secretary of Health and Human Services, not later than 36 months after Nov. 28, 1990, to prepare and submit to appropriate committees of Congress a report containing an evaluation of the requirements of 21 U.S.C. 360i(b), consisting of an evaluation of the safety benefits of the requirements, the burdens placed on the Food and Drug Administration and on device user facilities by the requirements, and the cost-effectiveness of the requirements and recommendations for legislative reform.

Section Referred to in Other Sections

This section is referred to in sections 331, 333, 352, 360c, 360e, 360g, 360j, 374 of this title.